- 19 -

1-Point Calibration:

CAL

ENT

ENT

CAL

HOLD

CAL

pH

READY

pH

HOLD

CAL

°C

pH

mV

HOLD

MEAS

°C

READY

ATC

pH

11

From pH measurement mode press CAL key to enter calibration mode
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to
begin calibration.

2

Place the electrode in pH 6.86 buffer. Immerse the temperature probe in
the buffer solution if ATC mode is enabled. Immerse the potential
matching pin in the buffer if symmetrical mode is enabled. The buffer

annunciator

appears in LCD. Lower display shows pH 6.86 (NIST

standard buffer). Upper display shows the current uncalibrated pH
reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator
when the reading is stable. Press CAL key to confirm the reading.

3

The calibration is completed. The pH Transmitter re-calculates electrode
properties based on the calibration. The new slope (in mV) is shown in the
upper display and normalized temperature is shown in the lower display.
Press ENT key to exit from the calibration.

4

The pH Transmitter reverts to pH measurement mode.

2-Point Calibration:

CAL

CAL

pH

READY

pH

CAL

pH

READY

pH

CAL

CAL

pH

READY

pH

ENT

ENT

ENT

ENT

CAL

pH

mV

MEAS

°C

READY

ATC

pH

ENT

11

From pH measurement mode press CAL key to enter calibration mode as
described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to
begin first calibration point.

2

Place the electrode in pH 6.86 buffer. Immerse the temperature probe in
the buffer solution if ATC mode is enabled. Immerse the potential matching

1

4

1

2

3

1

1

2

3

4

5

5

3