5

S E T U P M O D E

5.1

Enter Setup mode

In the Setup mode the transmitter can be configured to the individual requirements.

MEAS

°C

READY

ATC

pH

“User-Setting” Mode

(User can only

view parameters)

ENT

ENT

ENT

“User-setting” Mode

(User can change

parameter settings)

1. While in measurement mode, press the ENT key.

2. The display prompts to enter the security code. Set the security code with ▲ or

▼ key to:

“SCD 22” to change parameter settings

“SCD 0” to view parameter settings only (view only mode)

3. Press

the

ENT key.

NOTE:

To exit from any intermediate steps, press ▲ and ▼ keys (Escape)
together. It will return to measurement mode.