2.3 Connecting Peripherals

2.3.1

Connection Terminals

Remove Back Cover:

• Remove the screws from the four corners at the back of the pH

Transmitter.

• Remove the back cover. The connectors are exposed on the back

PCBA as shown in the Figure 1 below.

Connectors:

J202 – 9V DC power

• J204 - pH electrode & Temperature probe connections (wiring has to

be done in the detachable connector:

1. Pin 1

: pH Sensor

2. Pin 2

: pH Reference

3. Pin 3

: No Connection

4. Pin 4

: No Connection

5. Temp I/P

: Pt 100 Compensate

6. Temp Sens

: Pt 100 Sense

7. Temp Gnd

: Pt 100 GND

8. PMP

: Potential Matching Pin

Figure 1: Outer Side of Back PCBA