2 temperature calibration, Temperature calibration – Thermo Fisher Scientific Eutech pH 700/Ion 700 User Manual

Page 8

background image

5

4) To calibrate another buffer repeat step 3) or press

to return to the

measurement mode.

Note: The meter will automatically return to

measurement mode upon successful completion of the number of

specified calibration points. To specify a different number of pH

calibration points see

Section 5.6.

Additional notes:

A single point (offset) calibration is only allowed with pH 7.00 or

pH 6.86 buffers.

When the first calibration value is accepted during a new calibration,

all prior calibration values are erased.


Press

at any time to abort calibration and return to measurement

mode.


3.2 Temperature Calibration
The thermistor sensor used for automatic temperature compensation and

measurement is both accurate and stable, so require frequent calibration

isn’t required. Temperature calibration is recommended upon electrode

replacement, whenever the temperature reading is suspect, or if matching

against a certified thermometer is desired.

1) Connect the temperature probe to the meter and place into a solution

with a known accurate temperature such as a constant temperature

bath.

Note: To adjust the manual temperature compensation (MTC)

value, do not connect the temperature probe.

2) Press

as needed to select pH or mV/R.mV.

3) Press

followed by

. The primary display shows the

measured temperature while the secondary display shows the factory

default temperature.

4) Adjust the temperature using

or

. Press

to accept or

to cancel. The meter allows an adjustable maximum value of ± 5ºC

(or ± 9ºF) from the factory default temperature.