PROBLEM

CAUSE

SOLUTION

From any screen

Invalid key;

Button does not

work in the current

operation mode.

Press alternate key.

Select valid buttons

depending on mode.

Ion CAL mode

Ion slope calibration error;

calibration solutions are

outside acceptable range

(15 to 90 mV per decade).

Recalibrate using fresh ion

solutions. Use required ISA.

Replace electrode.

Ion CAL mode

Calibration points

are more than 1

decade apart.

Recalibrate using

consecutive calibration

values 1 decade apart

(ex. 1.0 and 10.0).

Ion CAL mode

Calibration attempted

with only 1-point during

ion calibration.

Recalibrate using 2 or more

ion calibration points.

pH CAL mode

Calibration attempted

using only 1-point that

was not pH 7 (USA) or

pH 6.86 (NIST).

Recalibrate using 2 or more

pH points or use pH 7 (USA)

or use pH 6.86 (NIST).

Err with electrode

icon

Calibration error.

Buffer value does not

match value displayed or

electrode is disconnected

or failing.

Use fresh buffer solutions.

Check electrode connection.

Clean & recondition

electrode.

Replace electrode.

Ur (Under range)
Or (Over range)

Measured value is out

of range.
Electrodes not connected.
Electrode clogged, dirty

or broken.
Meter not calibrated.
Wrong temp value.

Make sure electrode is

connected.
Clean or replace electrode.
Recalibrate the meter.
Treat samples to bring

within meter measuring

range.

Reset meter.