7. TROUBLESHOOTING GUIDE

PROBLEM

CAUSE

SOLUTION

No display

a) Main power not

switched on.

b) AC Adapter

socket not

inserted

properly.

a) Switch on the power supply.

b) Re-insert AC Adapter & press

.

Unstable

reading

a) Insufficient

reference

electrolyte in

electrode.

a) Fill electrode with reference electrolyte.

b) Broken

electrode.

b) Replace electrode.

c) External ‘noises’

or induction

(e.g. electrical

‘noise’ caused

by a nearby

motor).

c) Remove or switch off interfering device.

d) Dirty electrode.

d) Clean & rehydrate the electrode.

Slow

response

a) Dirty electrode.

a) Clean & rehydrate the electrode.

Meter not

responding

to key press

a) HOLD mode

in operation.

b) Bad keypad.

a) Press

to deactivate. See SETUP

to disable Auto Hold feature if enabled.

b) Contact Technical Service.