9

Note: When multiple slopes exist, the slope that is available to view will

depend on the measured value before entering SETUP mode.

For example, if pH calibration values pH 4.01, 7.00 and 10.01 were

completed, there will be two slope values; one that corresponds to pH

values below 7.00 and another that corresponds to pH values above 7.00.

If the measured pH value was 5.23 when setup mode was entered, the pH

slope that corresponds to values lower than 7.00 will be available to view.

If no calibration is stored, the offset will be 0.0 mV and the slope is 100%.


5.3 P3.0 ConF (Configuration)

Press

.


5.4 P3.1 rdY (Ready / Stability Indicator)

Press

.

Press

or

to choose READY “On”, READY “OFF”, or

Auto HOLd.
Press

to confirm.


5.5 P3.2 ºCºF (Celcius Or Fahrenheit)

Press

.

Press

or

to select ºC or ºF.

Press

to confirm.

Last Display

Mode

Electrode Property

Offset

Slope

pH

in mV

%

Ion

-

in mV

mV

in mV

-