Thermo Fisher Scientific Eutech pH 700/Ion 700 User Manual

Page 10

background image

7

1) Connect the ISE and press

as needed to select ppm.

Note: The

primary reading will show “- - - -” without a stored ion calibration. The

secondary value is the corresponding mV reading of the ISE.

2) Dip the ISE into your lowest standard solution value and stir. Press

when the secondary reading is stable.

3) Press

or

to match the primary display to your corresponding

ion calibration value (0.10, 1.0, 10, etc).

4) Press

to accept or

to abort. The primary display will show the

next highest calibration standard value.

5) Rinse the ISE then dip into the corresponding calibration standard

and stir.

6) Press

to accept or

to cancel.

The mV/decade slope value will

be displayed briefly if the calibration is successful. “SLP Err” indicates

that the calibration for the current point was not successful. This occurs

when the slope (mV difference between two consecutive points) is

lower than 15mV/decade or higher than 90mV/decade.

7) Revert to step 3) to calibrate additional points or press

as needed

to return to measurement mode at any time.

Note: The meter will

automatically return to measurement mode upon successful completion

of the number of specified calibration points. To change the number of

ion calibration points see

Section 5.6.