16

9. SPECIFICATIONS

pH Range

-2.00 to 16.00 pH

Resolution

0.01 pH

Accuracy

±

0.01 pH

Calibration Points

Up to 5 points with Auto-buffer recognition

Buffer Options

USA : pH 1.68, 4.01, 7.00, 10.01,12.45

NIST: pH 1.68, 4.01, 6.86, 9.18 ,12.45

Slope Display

Yes (with offset)

mV Range

±

2000 mV

Resolution

0.1 mV (± 199.9 mV), 1 mV beyond ± 200 mV

Accuracy

±

0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200 mV

Offset Adjustment

Up to ±150 mV

Temperature Range

0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution

0.1 ºC / ºF

Accuracy

±

0.3 ºC/± 0.5 ºF (0 to 70 ºC)

Compensation

Automatic or Manual (0 to 100 ºC / 32 to 212 ºF)

Calibration

Offset in 0.1 º increments; Offset range: ± 5.0 ºC / 9.0 ºF

Ion Range (Ion 700

only)

0.01 to 2000 ppm ( ± 2000mV)

Resolution (ppm)

0.01/ 0.1/1 ppm

Accuracy

±0.5% FS (mono-valent) ±1.0% Full Scale

(di-valent)

Calibration Points

From 2 to 5 consecutive points;

0.1, 1.0, 10.0, 100.0 or 1000.0 ppm

Slope Range

15 mV/decade to 90 mV/decade

Memory

100 data sets

Inputs

BNC, ATC, Reference (Half-cell)

Power

AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)