Thermo Fisher Scientific CyberScan CON 400/410 (For units manufactured before March 2010, discontinued) User Manual

Page 43

background image

Instruction Manual

CON 400/410

- 37 -

7.8

P7.0: Setting the real-time clock

The meter features a real-time calendar and clock. This
helps you to meet the Good Laboratory Practice (GLP)
standards.

From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P7.0.

3. Press

the

ENTER key to enter parameter P7.0. The

meter lets you select the century: “19—“ or “20—“. The
century digits will flash.

4. Press

the

MI/S or MR/T keys to toggle to the correct

century.

5. Press

ENTER key to confirm the century and move to

“year” selection. The “year” digits will flash.

6. Press

the

MI/S or MR/T keys to toggle to the correct

year. See Figure 33.

7. Press

ENTER key to confirm the year and move to

“month” selection. The “month” digits will flash.

8. Press

the

MI/S or MR/T keys to toggle to the correct

month.

9. Press

the

ENTER key to confirm the month and move

to “date” selection. The “date” digits will flash. See
Figure 33.

10. Press

the

MI/S or MR/T keys to toggle to the correct

date.

11. Press

the

ENTER key to confirm the date and move to

“hour” selection. The “hour” digits will flash. See Figure
34 (next page).

Figure 33 - P5.0: Setting century,

year, month and date