7.6.2
P5.2 Adjusting the normalization temperature
Your meter will normalize its conductivity measurements to a standard temperature that you
can select. You can adjust the normalization temperature from 15 to 30
°C (59 to 86 °F). Meter
default is 25
°C (77 °F).
From measurement mode
1. Press
SETUP key to enter Set Up mode.
2. Press
the
MI/S or MR/T keys to scroll through
subgroups until you view parameter P5.0.
3. Press
the
ENTER key three times to select
parameter 5.2. The display shows “t.nr” on the
upper display.
4. Press
the
ENTER key again. The upper display
shows the normalisation temperature and the lower
display shows “t.nr”.
5. Press
the
MI/S or MR/T keys to select the
normalisation temperature.
6. Press
the
ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.
NOTE:
Most calibration standards are referenced to 25°C. If a
normalization temperature other than 25°C is entered, the
calibration standard value at this temperature must be
known.
Figure 31 - Adjusting the
normalization temperature