P5.2 Adjusting the normalization temperature

Your meter will normalize its conductivity measurements to a standard temperature that you
can select. You can adjust the normalization temperature from 15 to 30

°C (59 to 86 °F). Meter

default is 25

°C (77 °F).


From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P5.0.

3. Press

the

ENTER key three times to select

parameter 5.2. The display shows “t.nr” on the
upper display.

4. Press

the

ENTER key again. The upper display

shows the normalisation temperature and the lower
display shows “t.nr”.

5. Press

the

MI/S or MR/T keys to select the

normalisation temperature.

6. Press

the

ENTER key to confirm selection and to

return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.



NOTE:
Most calibration standards are referenced to 25°C. If a
normalization temperature other than 25°C is entered, the
calibration standard value at this temperature must be
known.

Figure 31 - Adjusting the

normalization temperature