Instruction Manual

CON 400/410

- 34 -

7.6 P5.0

Temperature

7.6.1

P5.1 Adjusting the temperature coefficient

The temperature coefficient is the amount of change in conductivity per degree of temperature;
it is expressed in percent per

°C. Entering the exact temperature coefficient of your solution

lets you accurately compensate temperature for almost any solution. You can adjust 0.0 to
10.0 % per

°C. Meter default is 2.1% per °C. Adjustment to 0.0% will not correct for

temperature effects so the displayed value will be the actual value at the displayed
temperature.

From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P5.0.

3. Press

the

ENTER key to select parameter 5.1. The

display shows “t.CO” on the upper display.

4. Press

the

ENTER key again. The upper display

shows the temperature coefficient and the lower
display shows “t.CO”.

5. Press

the

MI/S or MR/T keys to select the

temperature coefficient of your solution.

6. Press

the

ENTER key to confirm selection and to

proceed to step 3 of P5.2. Press the CAL/MEAS key
twice to return to measurement mode.


NOTES:
If you do not know the temperature coefficient of your
solution you can determine the correct value using the
formula in Addendum 3 “Calculating Temperature
Coefficients” on page 47.

Figure 30 - Changing the

temperature coefficient