Instruction Manual

CON 400/410

- 27 -

7.2

P 1.0: Memory Clear (CLr)

Use this parameter to clear all stored memory values when you need to store a new series of
values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.

NOTE: Selecting YES will wipe out all memory.

From measurement mode:
1. Press

the

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view the parameter P1.0.

3. Press

the

ENTER key to enter parameter P1.0.

4. Press

the

MI/S or MR/T keys to toggle between

NO and YES. NO retains current memory; YES
clears all memory. See Figure 23.

5. Press

the

ENTER key to confirm selection and

return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.


Figure 23 - P1.0: Memory Clear