Instruction Manual

CON 400/410

- 32 -

7.5.3

P4.3 Selecting Automatic or Manual Temperature Compensation

This feature lets you select between Automatic Temperature Compensation (ATC) and
Manual Temperature Compensation. Meter default is ATC activated.

From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P4.0.

3. Press

the

ENTER key three times to select parameter

4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.

4. Press

the

ENTER key again. The upper display

shows “ATC” and the lower display shows “YES” or
“NO”.

5. Press

MI/S or MR/T keys to select Automatic or

Manual Temperature Compensation.
• YES = Automatic Temperature Compensation

activated.

• NO = Manual Temperature Compensation

activated.

6. Press

the

ENTER key to confirm selection and to

return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.

Figure 28 - Selecting Automatic or