Tension Measurement

To measure tensile force, use the hook attachment. The

display will show the force measured and a minus sign (-)

will appear on the left of the display (to the left of the

digits).

NOTE: To display no sign (plus) for tensile force, consult

the “Function Setting” section.

Compression Measurement

To measure compression force, use the flat head

attachment. No sign will appear on the left of the display

during a compression measurement.

NOTE: To display minus sign (-) for compression force,

consult the “Function Setting” section.

Low Battery

When battery charge is low, LO BAT will appear on the

display indicating the batteries need to be charged.

Charging time of fully depleted batteries is approximately

18 hours when the unit is off. The adapter/charger

automatically shuts off when the battery is at full charge

to protect the battery.

NOTE: The adapter/charger can be used to power the unit

during battery charging, but will lengthen charging time.

Auto Power Off

If the gauge is on and there is no activity for 10 minutes,

the unit automatically powers off to conserve battery

charge. PWR appears above the display digits to notify

that there is 1 minute before power off. If the adapter/

charger is powering the gauge, auto power off function

becomes inactive.

NOTE: To modify the auto power off function, consult

the “Function Setting” section.

External Device Communications

When operated with the appropriate output cable, the

FGV-HX offers output capabilities.

Overload

Caution should be exercised that overloads do not occur;

sensor damage can result. To protect the gauge and

sample under test when used with a motorized stand

(or similar device), two built-in OC NPN transistors stop

motion when the gauge reaches 120% of its rated

capacity. One transistor is for tension, the other for

compression (figure 2).

Analog Output

An analog output signal is available for recording

purposes. The amplitude of this signal is ± 1 VDC. The

voltage is positive when compression testing is performed

and negative for tension. The signal characteristics are

as follows:

Amplitude

Amplitude

Amplitude

Amplitude

Amplitude

± 1 VDC

Generated by

Generated by

Generated by

Generated by

Generated by

12-bit D/A converter

Signal update

Signal update

Signal update

Signal update

Signal update

100 times/sec (update every 10 msec)

Load impedance

Load impedance

Load impedance

Load impedance

Load impedance

10 K

minimum

Connector pins

Connector pins

Connector pins

Connector pins

Connector pins

Pin #1

signal output (analog)

Pin #2

GND (analog)

(see Fig.3 & Table 1)

NOTE: When the zero switch is pressed to tare the gauge

the analog output goes to 0V automatically.

Figure 1

Figure 2