Flash tube replacement, Specifications – SHIMPO DT-721 User Manual
Page 3
When FPM reading is displayed but unit is not flashing,
flash tube may need to be replaced.
1. Unplug line cord from power line. Turn power switch
OFF. Wait a few minutes until stroboscope is cool
before replacing flash tube.
2. Remove protective window by removing the 4 screws.
3. Use a rag and pull the tube out of its socket by rocking
it slightly up and down. Do not use bare hands to
remove tube, it may break and cause injury.
4. Insert new tube using the technique mentioned above.
Make sure that the tube is placed properly in the socket
otherwise it will touch the reflector. Tube should be set
symetrically within the neck of the reflector.
5. Replace protective window.
Flash Tube Replacement
MODEL
MODEL
MODEL
MODEL
MODEL
DT
DT
DT
DT
DT- 721(AC)
- 721(AC)
- 721(AC)
- 721(AC)
- 721(AC)
FLASHING RANGE
FLASHING RANGE
FLASHING RANGE
FLASHING RANGE
FLASHING RANGE
40 — 12,500 FPM (Flashes Per Minute)
ACCURACY
ACCURACY
ACCURACY
ACCURACY
ACCURACY
±0.02% of reading
RESOL
RESOL
RESOL
RESOL
RESOLUTION
UTION
UTION
UTION
UTION
0.1 ,
40.0 – 4,999.9 FPM
0.2 , 5,000.0 – 7,999.8 FPM
0.5 , 8,000.0 – 9,999.5 FPM
1
.0
, 10,000
.0
– 12,500
.0
FPM
DISPL
DISPL
DISPL
DISPL
DISPLAAAAAYYYYY
5-Digit LED, 0.3" (8 mm) height
R
RR
RRAAAAATE
TE
TE
TE
TE
Divide by 2, Multiply by 2
UPDA
UPDA
UPDA
UPDA
UPDATE TIME
TE TIME
TE TIME
TE TIME
TE TIME
Ext. mode: varies with flashing rate
OPER
OPER
OPER
OPER
OPERAAAAATING TIME
TING TIME
TING TIME
TING TIME
TING TIME
Continuous
FLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFE
FLASH TUBE POWER/LIFE
Xenon, 10 W, 100 million flashes
FL
FL
FL
FL
FLASH DUR
ASH DUR
ASH DUR
ASH DUR
ASH DUR AAAAATION
TION
TION
TION
TION
10 - 15
µ
s
SIGNAL OUTPUT
SIGNAL OUTPUT
SIGNAL OUTPUT
SIGNAL OUTPUT
SIGNAL OUTPUT
NPN Open Collector (24 VDC max.,
50 mA) 150
µ
s typical
PHASE SHIF
PHASE SHIF
PHASE SHIF
PHASE SHIF
PHASE SHIFTTTTT
Internal mode only by using the “+”
and “–” switches
EXT
EXT
EXT
EXT
EXT. TRIGGER
. TRIGGER
. TRIGGER
. TRIGGER
. TRIGGER
a) 12-24 VDC thru a 3-wire NPN output
sensor (requires external power supply)
INPUT SIGNAL
INPUT SIGNAL
INPUT SIGNAL
INPUT SIGNAL
INPUT SIGNAL
b) 12 VDC thru a 2-wire proximity
sensor (leakage current 1 mA max.,
Load current 8 mA min.)
c) Switch or Relay contact
POWER REQUIREMENTS
POWER REQUIREMENTS
POWER REQUIREMENTS
POWER REQUIREMENTS
POWER REQUIREMENTS
115 VAC ±10% 60/50 Hz
OPER
OPER
OPER
OPER
OPERAAAAATING TEMPER
TING TEMPER
TING TEMPER
TING TEMPER
TING TEMPERAAAAATURE
TURE
TURE
TURE
TURE
32° – 104° F (0 – 40° C)
DIMENSIONS
DIMENSIONS
DIMENSIONS
DIMENSIONS
DIMENSIONS
9" L x 4.3" W x 8.5" H
(Height includes handle)
WEIGHT
WEIGHT
WEIGHT
WEIGHT
WEIGHT
1.5 lbs. (0.7 Kg)
ACCESSORIES AVAILABLE
ACCESSORIES AVAILABLE
ACCESSORIES AVAILABLE
ACCESSORIES AVAILABLE
ACCESSORIES AVAILABLE
Carrying Case
Specifications