beautypg.com

Calibrator interface, Figure 1. input/output terminals – Martel Electronics MC1210 User Manual

Page 6

background image

2. Calibrator Interface

Figure 1 shows the location of the input and output connections on the calibrator, while Table
1 describes their use.

Table 1: Input and Output Terminals

No.

Name

Description

1, 2

Measure Isolated V,

Input terminals for measuring current, voltage, switchtest,

mA terminals, switchtest

and supplying loop power.

3

TC input/output

Terminal for measuring, or simulating thermocouples. Accepts
miniature polarized thermocouple plugs with flat in-line blades
spaced 7.9 mm (0.312 in) center to center.

4,5

Source/Measure

Terminals for sourcing and measuring voltage, frequency,

V,RTD 2W, Hz,

pulse train, and RTDs

6,7

Source/Measure

Terminals for sourcing and measuring current, and performing

mA terminals, 3W 4W

RTD measurements with 3-wire or 4-wire setups.

8

Pressure module connector

Connects calibrator to a pressure module for pressure meas-
urements.

9

Serial port

Connects calibrator to a PC for remote control.

6

5

3

4

2

1

9

8

7

6

F1

F2

F3

CE

ENTER

8

9

4

1

5

2

0

6

3

.

7

HOME

Figure 1. Input/Output Terminals