Martel Electronics MC1210 User Manual
Page 50
50
Table 12: Frequency Measurement/Source
Range
Accuracy (% of reading ± floor)
Read
2.0CPM - 600.0CPM
0.05% ± 0.1CPM
1.0Hz - 1000.0Hz
0.05% ± 0.1Hz
1.00KHz - 10.00KHz
0.05% ± 0.01KHz
Source
2.0CPM - 600.0CPM
0.05%
1.0Hz - 1000.0Hz
0.05%
1.00KHz - 10.00KHz
0.250%
Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is
adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight negative
offset of approximately -0.1V is present to assure zero crossing.
Table 13: Resistance Measurement
Range
Accuracy (% of reading ± floor)
Ohms low
0.00
Ω - 400.0Ω
0.025% ± 0.05
Ω
Ohms high
401.0
Ω - 4000.0Ω
0.025% ± 0.5
Ω
Table 14: Resistance Source
Range
Excitation Current
Accuracy (% of reading ± floor)
Ohms low
5.0
Ω - 400.0Ω
0.1mA - 0.5mA
0.025% ± 0.1
Ω
5.0
Ω - 400.0Ω
0.5mA - 3mA
0.025% ± 0.05
Ω
Ohms high
400
Ω - 1500Ω
0.05mA - 0.8mA
0.025% ± 0.5
Ω
1500
Ω - 4000Ω
0.05mA - 0.4mA
0.025% ± 0.5
Ω
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.
Table 15: Thermocouple Measurement/Source
Range
Accuracy (% of reading ± floor)
Read (mV)
-10.000mV - 75.000mV
0.02% ± 10
μV
Source (mV)
-10.000mV - 75.000mV
0.02% ± 10
μV
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1
Ω
Table 16: Thermocouple Read and Source (errors in °C)
Range (°C)
Accuracy (°C)
TC Type
Minimum
Maximum
CJC OFF
CJC ON
J
-210.0
-150.0
0.4
0.6
-150.0
1200.0
0.2
0.4
K
-200.0
-100.0
0.5
0.7
-100.0
600.0
0.2
0.4
600.0
1000.0
0.3
0.5
1000.0
1372.0
0.4
0.6