5 entering commands – Martel Electronics MC1210 User Manual

Page 39

background image

Units

Meaning

MMH2O4C

Pressure in millimeters of water at 4°C

MMH2O20C

Pressure in millimeters of water at 20°C

FTH2O60F

Pressure in feet of water at 60°F

Table 8: Error codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a numeric entry

101

Too many significant digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid pressure unit was received

107

An invalid CJC_STATE was received

108

An invalid TSENS_TYPE was received

109

Pressure module not connected

110

An unknown command was received

111

An invalid RTD or TC parameter value was received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

115

Output is overloaded

116

Calibrator not in pulse train mode when TRIG was received

117

An invalid FREQ_TYPE was received

8.5 Entering Commands

Commands for the calibrator may be entered in upper or lower case. There is at least one
space required between the command and parameter, all other spaces are optional. Almost
all commands for the calibrator are sequential, any overlapped commands will be indicated
as such. This section will briefly explain each of the commands and describe their general
use, which will include any parameters that may be entered with the command as well as
what the output of the command is.

8.5-1 Common Commands

*CLS

Clears the ESR, the error queue and the RQS bit. Also terminates all pending operations.
When writing programs, use before each procedure to avoid buffer overflow.

39