Instantaneous value measurement: w range, 1. steps for measurement – KYORITSU 6305 Quick Manual User Manual
Page 15

−
14 −
KEW6305
KEW6305
Instantaneous value Measurement:
W
Range
6. Instantaneous value Measurement:
W
Range
6-1. Steps for measurement
Ensuring your safety
↓
Preparation for measurement
↓
Setting
↓
Wiring
↓
Instantaneous value measurement
●
Parameters for
W
Range
Parameters to be displayed on the LCD
Unit
Voltage
(RMS)
V
: average voltage of each phase
Vi
: voltage of each phase
V
Current
(RMS)
A
: average current of each phase
Ai
: current of each phase
A
Active
power
P
: total active power
Pi : active power of each phase
Polarity :+ (no sign) consumption
- (minus) regenerating
W
Reactive
power
Q
: total reactive power
Qi : reactive power of each phase
Polarity :+ (no sign) lagging phase
- (minus) leading phase
Var
Apparent
power S
: total apparent power Si: apparent power of each phase
VA
Power
factor
PF
: power factor of whole system
PFi : power factor of each phase
Polarity :+ (no sign) lagging phase
- (minus) leading phase
PF
Frequency f
: frequency of V1
Hz
Neutral
current In
: current on neutral line (3P4W only)
An
* i = 1, 2, 3
SET UP Range
Basic setting
Measurement setting
“
Setting 01” Wiring system
“
Setting 02” Voltage range
“
Setting 03” Clamp sensor
“
Setting 04” Current range
“
Setting 05” VT ratio (if necessary)
“
Setting 06” CT ratio (if necessary)
Range