Set-up: set up range – KYORITSU 6305 Quick Manual User Manual
Page 10

−
9 −
KEW6305
KEW6305
Set-up:
SET UP
Range
4. Set-up:
SET UP
Range
Set the Function switch to
SET UP
Range for making various settings.
Settings listed below can be changed. (27 items in total)
List of setting items
Setting Setting no./ item
Symbol
Details
Basic
setting
01 Wiring system
1P2W(1ch)/ 1P2W(2ch)/ 1P2W(3ch)/
1P3W/ 3P3W/ 3P3W3A/ 3P4W
02 Voltage range
150/ 300/ 600V
03 Clamp sensor
50/ 100/ 200/ 500/ 1000/ 3000A type
04 Current range
-
03
Sensor Range
50A
1/5/10/25/50A/AUTO
100A 2/10/20/50/100A/AUTO
200A 4/20/40/100/200A/AUTO
500A 10/50/100/250/500A/AUTO
1000A 20/100/200/500/1000A/AUTO
3000A 300/1000/3000A
05 VT ratio
0.01 - 9999.99 (can be set by 0.01)
06 CT ratio
0.01 - 9999.99 (can be set by 0.01)
Time&
Buzzer
07 Date and time
Year:Month:Day:Hour:Minute:Second
08 Buzzer
ON / OFF
Measure-
ment
09 Recording interval
1/ 2/ 5/ 10/ 15/ 20/ 30 sec./
1/ 2/ 5/ 10/ 15/ 20/ 30 min./ 1 hour
10 Specific time period
rec. or endless rec.
ON: Specifying start/ stop time
(repeatedly recorded)
OFF : Record the data continuously
11
*1
Time period setting
Time setting
Start and stop time
(Year:Month:Day:Hour:Minute:Second)
12
*1
Time period setting
Date setting
Year:Month:Day:Hour:Minute:Second
13
*2
Start of continuous
measurement
Year:Month:Day:Hour:Minute:Second
14
*2
End of continuous
measurement
Year:Month:Day:Hour:Minute:Second
15 Target demand
Value : 0.1 - 999.9
Unit: W/kW/MW/GW/VA/kVA/MVA/GVA
16 Demand measure-
ment cycle
NO/ 10/ 15/ 30 min
* Demand measurement will not be
performed when “NO” has been
selected.
17 Demand warning
cycle
1/2/5 min. when measurement cycle is
10or15 min., 1/2/5/10/15 min. when
measurement cycle is 30 min.
*1 : Setting 11& 12 can be changed only when Setting 10 has been set to “ON”.
*2 : Setting 13& 14 can be changed only when Setting 10 has been set to “OFF”.