KYORITSU 3128 User Manual
Page 4
P. 49 Additional sentences at
“6.4.2 How to measure PI”
DAR Simultaneous display
DAR value is displayed during PI measurements and the measured results are being displayed.
TIME 1 and 2 values for DAR are the ones pre-set in the DAR mode. Refer to 6.5.2 How to measure DAR
and set the time for DAR.
In case that DAR TIME 2 value is higher than PI TIME 2 value, DAR value isn
’t displayed on the LCD. A
measurement stop when the PI TIME 2 comes. PI value isn
’t displayed in the DAR measurement mode.
P. 52 Amendments to the illustration and table at
“6.4.3 Measured Result”
Displayed Items
Details
Insulation Resistance
Measured insulation resistance value
PI Resistance at TIME1,2
PI Resistance value at TIME1 and TIME2
DAR Resistance at TIME1,2
DAR Resistance value at TIME1 and TIME2
PI
Polarization Index value
Pre-set Voltage
Pre-set output voltage value
Output Voltage
Voltage being output
Current value
Current value being measured
Capacitance
Capacitance measured at discharge
DAR
Dielectric Absorption Ratio
P. 61 Additional sentence at
“6.7.2 Measured Setting Items”
In the SV measurement mode, measurements continue after the preset Step time (V5) comes and
automatically stop when 90 min pass.
P.70 Amendments to the description at
“6.11.3 Auto-power-off Function”.
The instrument automatically turns off when approx 10 min pass after a measurement with Timer function
activated or 90 min of continuous measurement in SV mode completes.
PI value
DAR Resistance
at TIME1,2
PI Resistance
at TIME1,2
電圧設定値
Pre-set Voltage
出力電圧値
Output Voltage
Capacitance
DAR value
PI値
Current value
絶縁抵抗値
Insulation