Measurement time, Condition current consumption – KYORITSU 3128 User Manual
Page 2
P.11 Additional graph of
“Resistance-Output voltage Characteristic” for each Voltage range
[Graph of Resistance-Output voltage Characteristic]
Rated current at 500V, 1000V, 2500V, 5000V Ranges : max 1mA or higher
Rated current at 10kV, 12kV Ranges : max 0.5mA
Above graphs show the relations between output voltages and measurement resistances.
*
“Rated measurement current” described in the table at page 11 should be corrected to “Output current”.
P.15 Additional description about current consumption and battery life
Continuous measurement: No limitation (IR measurement mode)
* recorded data and graph are max 90 min.
Max 90 min (SV measurement mode)
Max 60 min (PI/DAR/DD measurement mode)
Max current consumption and measurement time while rated currents are maintained.
Condition
Current consumption
Measurement time
500V /
300kΩ
2100mA
or less
Approx
2 hours
1000V / 600kΩ
2500V / 2.4MΩ
5000V / 4.8MΩ
10000V / 20MΩ
12000V / 24MΩ
* Current consumption may exceed above values when low resistances, which influence the output of
rated currents, are measured.
* Measurement time (approx 4 hours) described at page 15 and the Measurement time in above table
are the periods that full-charged battery voltage drops to the lowest level.
* It is recommended to charge the battery with reference to
“7.1 How to charge battery” described in the
manual before stating to use with the instrument since the battery voltage may be low due to
self-discharge.
10kV/12kV Range
0
2,000
4,000
6,000
8,000
10,000
12,000
14,000
0.0001
0.001
0.01
0.1
1
10
100
Insulation Resistance (G
Ω )
O
u
tp
u
t
Vo
lt
ag
e
(V)
12kV
10kV
500V/1000V/2500V/5000V Range
0
1,000
2,000
3,000
4,000
5,000
6,000
0.0001
0.001
0.01
0.1
1
10
100
Insulation Resistance (G
Ω )
O
u
tp
ut
V
olt
ag
e
(V
)
500V
1000V
2500V
5000V