KYORITSU 3128 User Manual
Page 14
10
KEW3128
3. Specification
KEW3128
●
Measuring Range and Accuracy (under 23±5ºC and 45 – 75%RH)
【
Insulation Resistance Tester】
Rated Voltage
500V
1000V
Max value
500GΩ
1TΩ
Accuracy
0~50GΩ
±5%±3dgt
* Accuracy
is not
guaranteed
with setting
of 250V
or less.
0~100GΩ
±5%±3dgt
50G~500GΩ
±20%
100G~1TΩ±20%
Short-circuit
current
Max 5.0mA
Rated
measurement
current
1mA or more, 1.2mA
or less under a load of
0.5MΩ
* Should be 500V or more
1mA or more, 1.2mA
or less under a load of
1MΩ
Rated Voltage
2500V
5000V
Max value
2.5TΩ
5TΩ
Accuracy
0~250GΩ
±5%±3dgt
0~500GΩ
±5%±3dgt
250G~2.5TΩ±20%
500G~5TΩ±20%
Short-circuit
current
Max 5.0mA
Rated
measurement
current
1mA or more, 1.2mA or
less under a load of
2.5MΩ
1mA or more, 1.2mA or
less under a load
of 5MΩ
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)