KEW3128
25
KEW3128
6. Measurement
⑧
The measured value will be displayed on the LCD.
⑨ Release the Button to stop the measurement. Then the measured
results will be displayed on the LCD.
Turn the locked Test Button counterclockwise and unlock the Button.
⑩ This instrument has an auto-discharge function. With the test
leads connected to the circuit under test, release the Test button
to discharge capacitance in the circuit after test.
Confirm that the readings on the Voltage monitor becomes “0V”.
* When the test leads are disconnected from the measured object
during a discharge, keep the same connection and connect the
test leads to the measured object again and continue a discharge.
Attention should be paid when connecting the test leads and the
measured object again so as not to cause aerial discharges between
them for a long period. In this case, a longer discharge period is
required because the internal discharge circuit doesn’t work.
⑪ Set the Range Switch to“OFF”position, and remove the test
leads from the device under test.
Next measurement may not be started when the Test Button is pressed
right after when above step 10 is done. In this case, release the Test
Button and wait for a few seconds, and then press the Test Button.
Refer to “
IR Measurement” (=>P.48) and subsequent pages about
the further details of items displayed at each Measurement mode.