beautypg.com

KYORITSU 3128 User Manual

Page 17

background image

KEW3128

13

KEW3128

3. Specification

● Operating system

Dual integration

● Display

320 x 240 dots, 5.7-inch

Monochrome Display

● Low battery warning

Battery mark display (in 4 levels)

● Response time

approx 30 sec in a range of ±5% of accuracy

approx 60 sec in a range of ±20% of

accuracy (Response time becomes slower

when output voltage becomes lower.)

● Auto-power-off

Power-off function operates when 10 min

pass without any key operation.

● Altitude

2000m or less

● Temperature & humidity

range(guaranteed accuracy)

23ºC±5ºC/Relative humidity 85% or less

(no condensation)

● Operating temperature &

humidity range

-10ºC~50ºC/Relative humidity 85% or less

(when operating with an external power

supply, no condensation)

0ºC~40ºC/Relative humidity 85% or less

(when operating with battery, no condensation)

● Storage temperature &

humidity range

-20ºC~60ºC/Relative humidity 75% or less

(no condensation)

● Overload protection

AC720V/10 sec.

● Withstand voltage

AC8770V : between line terminal and

enclosure /5sec (50/60Hz)

AC6880V : between the measuring

terminal and enclosure /5sec (50/60Hz)

AC2330V : between the power connector

and enclosure /5sec (50/60Hz)

● Insulation resistance

1000MΩ or more/DC1000V

(between electrical circuit and enclosure)

● Dimension

330(L)×410(W)×180(D)mm

(Instrument and Hard case )

● Weight

approx 9kg (including battery)

(Instrument and Hard case )

● Power source

Rechargeable Lead storage battery

( PXL-12050:12V 5Ah), AC Power supply

(100V ~ 240V, 50 / 60Hz)