KYORITSU 3127 User Manual
Page 37

- 34 -
7. Memory Function
7-1 Function Details
Measurement data of insulation resistance can be saved in the internal memory
of KEW3127.
Following data can be saved.
1. LOGGING: Measuring data is saved every second.
2. MEMORY: Data is saved at the end of measurement.
(1) Max number of file
1. LOGGING: Max 10 files
* Up to 100min in total
* Max recording time per file: 90min
2. MEMORY: Max 32 files
(2) Parameters to be saved
Following parameters will be saved. (common for LOGGING and MEMORY)
1. In all modes: Saved time & date, measured values (resistance, current,
voltage), capacitance
2. IR
PI/DAR
mode: PI/ DAR value
DD mode: DD value
Ramp mode: Breakdown voltage
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)