beautypg.com

Technical data, Temperature range, Hardrubber – KROHNE OPTIFLUX 2070 EN User Manual

Page 32: Polypropylene

background image

8

TECHNICAL DATA

32

OPTIFLUX 2070

www.krohne.com

04/2008 • 4000243601

Temperature range

Process [°C]

Ambient [°C]

min.

max.

min.

max.

Hardrubber

Separate flow sensor (OPTIFLUX

2000 F)

-5

80

-40

65

Compact with IFC 070 (OPTIFLUX

2070 C)

-5

80

-40

65

Polypropylene

1

Separate flow sensor (OPTIFLUX

2000 F)

-5

90

-40

65

Compact with IFC 070 (OPTIFLUX

2070 C)

-5

90

-40

65

1

Polypropylene available for DN 50 - 150

Process [°F]

Ambient [°F]

min.

max.

min.

max.

Hardrubber

Separate flow sensor (OPTIFLUX

2000 F)

23

176

-40

149

Compact with IFC 070 (OPTIFLUX

2070 C)

23

176

-40

149

Polypropylene

1

Separate flow sensor (OPTIFLUX

2000 F)

23

194

-40

149

Compact with IFC 070 (OPTIFLUX

2070 C)

23

194

-40

149

1

Polypropylene available for ASME 2" - 6"