SERVICE

7

27

OPTIFLUX 2070

www.krohne.com

04/2008 • 4000243601

Replace the cover.

Tighten the 4 bolts.

Enter the programming mode, hold the V and Z key for 5 seconds.

i

The display flashes.

Press the Z key to enter the menu.

Scroll through the available positions with the V key to position 74 and check the battery

capacity (important for battery life indication).

Go to position 75 and enter "1" (reset battery counter).

i

Battery capacity symbol at the display should be "full".

Hold the V key for 3 seconds to confirm the new value and go back to the measuring mode.

Replace the blue protection cap.

WARNING!
Make sure that the battery cable is not jammed by the cover.