ELECTRICAL CONNECTIONS

4

19

OPTIFLUX 2070

www.krohne.com

04/2008 • 4000243601

4.4 Cable connection for field version

4.4.1 Flow sensor side

Prepare appropriate cable lengths (

1

...

3

).

Connect the shield to the U-clamp, the brown cable to terminal 7 and the white to terminal 8.

Connect the shield to terminal 1, the purple cable to terminal 2 and the blue to terminal 3.

Figure 4-3:

Cable connection at flow sensor side

1

cable length: 13 cm / 5"

2

cable length: 5 cm / 2"

3

cable length: 8 cm / 3"

4

brown + white cable, used for field current

5

purple + blue cable, used for electrode signals

6

shield (terminal 1 of connector X2 + U-clamp)