Operation – KROHNE OPTIFLUX 2070 EN User Manual
Page 24
6
OPERATION
24
OPTIFLUX 2070
www.krohne.com
04/2008 • 4000243601
Software version 2.0.0
Menu
No.
Description
Display
default
Selection list
Remarks
12
Measuring unit
x
0 = m3, 1 = USG
Customer defined
13
Fiscal metering 0
0 = No, 1 = Yes
Blocks menu
21
Meter size
xxx
Defined at factory callibration -
22
Meter constant
xx.xxx
Defined at factory callibration
-
23
Zero offset
calibration
0
Set to 1
In situ determination of zero
point.
Confirm with “>” for 3 seconds
After countdown meter switches
to measuring mode.
Menu no. 24 is automatically set
to 1.
24
Zero selection
0
0 = factory calibration
Selection which zero point to use
1 = measured
25
Flow direction
0
0 = Forward, 1 = Reverse
-
26
Measuring rate
10
1, 5, 10, 15, 20 seconds
Measurement interval in seconds
27
Low flow cut off 20
0, 5, 10, 20 mm/s
Below value no measurement
28
Time constant
flow reading
2
1 = fast, 2 = normal, 3 = slow
Timeconstant of display
29
Factory
calibration /
comm. mode
0
0 = No, 1 = Yes
For factory use only, blocks
menu
30
Self check
0
0 = off, 1 = on
-
32
Simulate
outputs
0
0 = off, 1 = On
1 puls / second (for cable testing)
41
Output A
(pulse)
0
0 = off, 1 = on
-
42
Output B
(pulse)
0
0 = off, 1 = on
-
43
Phase shift
pulse output
90
90, 180 degrees offset or
F - r ( A-forward, B-reverse flow)
44
Pulse width
1
1, 5, 10, 50, 100 ms
-
45
Pulse value
xx.xxx
in m3 / pulse or USG / pulse
-
51
Status output 1
Self checking
0
0 = off, 1 = on
Active at instrument failure
52
Status output 1
Battery pre
warning
0
0 = off, 1 = on
Active 1 year before empty
battery
53
Status output 1
Battery final
warning
0
0 = off, 1 = on
Active at low battery
54
Status output 1
Counter
overrun
0
0 = off, 1 = on
After 99999999 counter starts at
zero
55
Status output 2
Self checking
0
0 = off, 1 = on
Active at instrument failure