6

OPERATION

24

OPTIFLUX 2070

www.krohne.com

04/2008 • 4000243601

Software version 2.0.0

Menu
No.

Description

Display
default

Selection list

Remarks

12

Measuring unit

x

0 = m3, 1 = USG

Customer defined

13

Fiscal metering 0

0 = No, 1 = Yes

Blocks menu

21

Meter size

xxx

Defined at factory callibration -

22

Meter constant

xx.xxx

Defined at factory callibration

-

23

Zero offset
calibration

0

Set to 1

In situ determination of zero
point.

Confirm with “>” for 3 seconds

After countdown meter switches
to measuring mode.

Menu no. 24 is automatically set
to 1.

24

Zero selection

0

0 = factory calibration

Selection which zero point to use

1 = measured

25

Flow direction

0

0 = Forward, 1 = Reverse

-

26

Measuring rate

10

1, 5, 10, 15, 20 seconds

Measurement interval in seconds

27

Low flow cut off 20

0, 5, 10, 20 mm/s

Below value no measurement

28

Time constant
flow reading

2

1 = fast, 2 = normal, 3 = slow

Timeconstant of display

29

Factory
calibration /
comm. mode

0

0 = No, 1 = Yes

For factory use only, blocks
menu

30

Self check

0

0 = off, 1 = on

-

32

Simulate
outputs

0

0 = off, 1 = On

1 puls / second (for cable testing)

41

Output A
(pulse)

0

0 = off, 1 = on

-

42

Output B
(pulse)

0

0 = off, 1 = on

-

43

Phase shift
pulse output

90

90, 180 degrees offset or

F - r ( A-forward, B-reverse flow)

44

Pulse width

1

1, 5, 10, 50, 100 ms

-

45

Pulse value

xx.xxx

in m3 / pulse or USG / pulse

-

51

Status output 1
Self checking

0

0 = off, 1 = on

Active at instrument failure

52

Status output 1
Battery pre
warning

0

0 = off, 1 = on

Active 1 year before empty
battery

53

Status output 1
Battery final
warning

0

0 = off, 1 = on

Active at low battery

54

Status output 1
Counter
overrun

0

0 = off, 1 = on

After 99999999 counter starts at
zero

55

Status output 2
Self checking

0

0 = off, 1 = on

Active at instrument failure