Technical data – KROHNE OPTIMASS 6000 EN User Manual
Page 58
6
TECHNICAL DATA
58
OPTIMASS 6000
www.krohne.com
07/2014 - 4001894003 - MA OPTIMASS 6000 R03 en
6.5.5 Purge port option
Dimensions
6.5.6 Burst discs
If the meter has been ordered with a burst disc, it will be supplied with the disc fitted. The failure
pressure of the disc is 10 barg at +20°C / 145 psig at +68°F.
mm
S08
S10
S15
S25
S50
S80
S100
S150
S200
S250
A
70
110
145
150
205
220
345
600
800
B
32
45
57
60
85
100
160
140
inches
S08
S10
S15
S25
S50
S80
S100
S150
S200
S250
A
2.7
4.3
5.7
5.9
8.1
8.7
13.6
23.6
31.5
B
1.3
1.8
2.2
2.4
3.3
3.9
6.3
5.5
CAUTION!
The fitted burst disc will be suitable for the flow rates and process conditions specified on the
original order. If the process conditions alter in any way, it is recommended that you contact the
manufacturer for advice regarding suitability.
If the process product is hazardous (in any way) it is recommended that an exhaust tube is
connected to the NPT male thread of the burst disc and the pipe routed so that the process
product can be discharged to a safe area. Use a tube with a diameter large enough AND routed in
such a way, so that pressure cannot build up in the meter case.
INFORMATION!
For gas applications, the burst disc must be specified at the time of ordering.
MA_OPTIMASS_6000_en_140717_4001894003_R03.book Page 58 Thursday, July 17, 2014 3:27 PM