6

TECHNICAL DATA

30

OPTIMASS 6000

www.krohne.com

07/2014 - 4001894003 - MA OPTIMASS 6000 R03 en

6.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Coriolis mass flow

Application range

Mass flow and density measurement of fluids, gases and solids

Measured values

Mass, density, temperature

Calculated values

Volume, referred density, concentration, velocity

Sensor model range

Sensor model range

Sensor model range

Sensor model range

Stainless Steel 316L 08...250

Compact / remote 100 barg @ 20°C / 1450 psig @ 68°F, temp. range

-70°C...+230°C / -94°F...+446°F

Remote only 100 barg @ 20°C / 1450 psig @ 68°F, temp. range

-50°C...+400°C / -58°F...+752°F

Remote only 100 barg @ 20°C / 1450 psig @ 68°F, temp. range

-200°C...+40°C / -364°F...+104°F

Hastelloy

®

08...80

Compact / remote 200 barg @ 20°C / 2900 psig @ 68°F, temp. range

-70°C...+230°C / -58°F...+446°F

Duplex Stainless Steel 100...200

Compact / remote 200 barg @ 20°C / 2900 psig @ 68°F, temp. range

-50°C...+230°C / -58°F...+446°F

Design

Basic

System consists of a measuring sensor and a converter to process the

output signal

Features

Fully welded maintenance free sensor with twin V-shaped measuring tube

Variants

Variants

Variants

Variants

Compact version

Integral converter

Remote version

Available with field or wall versions of the converter

Modbus version

Sensor with integral electronics. Modbus output for connection to a PLC

Measuring accuracy

Mass (standard)

Mass (standard)

Mass (standard)

Mass (standard)

Liquid (≥ 20:1 of nominal flow rate)

±

0.1% of actual measured flow rate

Liquid (< 20:1 of nominal flow rate)

±

zero stability (see zero stability below)

Gas

±

0.35% of actual measured flow rate + zero stability

Mass (optional)

Mass (optional)

Mass (optional)

Mass (optional)

Liquid (≥ 10:1 of nominal flow rate)

±

0.05% of actual measured flow rate

Liquid (< 10:1 of nominal flow rate)

±

zero stability (see zero stability below)

Repeatability

Repeatability

Repeatability

Repeatability

Liquid

Better than 0.05% plus zero stability (includes the combined effects of

repeatability, linearity and hysteresis)

Gas

Better than 0.2% plus zero stability (includes the combined effects of

repeatability, linearity and hysteresis)

MA_OPTIMASS_6000_en_140717_4001894003_R03.book Page 30 Thursday, July 17, 2014 3:27 PM