Technical data, 2 technical data – KROHNE OPTIMASS 6000 EN User Manual
Page 30
6
TECHNICAL DATA
30
OPTIMASS 6000
www.krohne.com
07/2014 - 4001894003 - MA OPTIMASS 6000 R03 en
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Coriolis mass flow
Application range
Mass flow and density measurement of fluids, gases and solids
Measured values
Mass, density, temperature
Calculated values
Volume, referred density, concentration, velocity
Sensor model range
Sensor model range
Sensor model range
Sensor model range
Stainless Steel 316L 08...250
Compact / remote 100 barg @ 20°C / 1450 psig @ 68°F, temp. range
-70°C...+230°C / -94°F...+446°F
Remote only 100 barg @ 20°C / 1450 psig @ 68°F, temp. range
-50°C...+400°C / -58°F...+752°F
Remote only 100 barg @ 20°C / 1450 psig @ 68°F, temp. range
-200°C...+40°C / -364°F...+104°F
Hastelloy
®
08...80
Compact / remote 200 barg @ 20°C / 2900 psig @ 68°F, temp. range
-70°C...+230°C / -58°F...+446°F
Duplex Stainless Steel 100...200
Compact / remote 200 barg @ 20°C / 2900 psig @ 68°F, temp. range
-50°C...+230°C / -58°F...+446°F
Design
Basic
System consists of a measuring sensor and a converter to process the
output signal
Features
Fully welded maintenance free sensor with twin V-shaped measuring tube
Variants
Variants
Variants
Variants
Compact version
Integral converter
Remote version
Available with field or wall versions of the converter
Modbus version
Sensor with integral electronics. Modbus output for connection to a PLC
Measuring accuracy
Mass (standard)
Mass (standard)
Mass (standard)
Mass (standard)
Liquid (≥ 20:1 of nominal flow rate)
±
0.1% of actual measured flow rate
Liquid (< 20:1 of nominal flow rate)
±
zero stability (see zero stability below)
Gas
±
0.35% of actual measured flow rate + zero stability
Mass (optional)
Mass (optional)
Mass (optional)
Mass (optional)
Liquid (≥ 10:1 of nominal flow rate)
±
0.05% of actual measured flow rate
Liquid (< 10:1 of nominal flow rate)
±
zero stability (see zero stability below)
Repeatability
Repeatability
Repeatability
Repeatability
Liquid
Better than 0.05% plus zero stability (includes the combined effects of
repeatability, linearity and hysteresis)
Gas
Better than 0.2% plus zero stability (includes the combined effects of
repeatability, linearity and hysteresis)