• Where liquid is being used, set up the pipe configuration so that air can be vented from the

system.

• Where steam is being used, set up the pipe configuration so that condensation can be drained

off.

• Gradually heat the jacket to working temperature before flowing the process fluid through

the meter.

3.4.12 Purge ports

If the meter has been ordered with a purge port, it will be supplied with NPT female connections
which are clearly marked. The connections are sealed with NPT plugs and PTFE tape.

3.4.13 Burst discs

If the meter has been ordered with a burst disc, it will be supplied with the disc fitted. The failure
pressure of the disc is 10 barg at +20°C / 145 psig at +68°F.

CAUTION!
The design of the heating jacket is such that NO end load pressure (negative or positive) can be
applied. The manufacturer recommends that (where possible) flexible hoses are used to connect
to the heat source.

CAUTION!
The maximum heating pressure and temperature for heating jackets is 10 barg at
230°C / 145 psig at 446°F or 5 barg at 400°C / 72.5 psig at 752°F

CAUTION!
DO NOT remove these plugs.

The meter is factory sealed with a dry nitrogen gas fill and if moisture is allowed to enter the
meter casing it will cause damage. The plugs should only be removed to purge the meter casing
in the event that the primary measuring tube fails.

If it is suspected that the primary measuring tube has failed, de-pressurise the meter and
remove it from service, as soon as it is safe to do so.

CAUTION!
The fitted burst disc will be suitable for the flow rates and process conditions specified on the
original order. If the process conditions alter in any way, it is recommended that you contact the
manufacturer for advice regarding suitability.

If the process product is hazardous (in any way) it is recommended that an exhaust tube is
connected to the NPT male thread of the burst disc and the pipe routed so that the process
product can be discharged to a safe area. Use a tube with a diameter large enough AND routed in
such a way, so that pressure cannot build up in the meter case.

INFORMATION!
For gas applications, the burst disc must be specified at the time of ordering.