beautypg.com

Technical data – KROHNE OPTITEMP Compact sensors User Manual

Page 34

background image

7

TECHNICAL DATA

34

OPTITEMP TRA-CXX

www.krohne.com

02/2014 - 4003290501 - MA OPTITEMP TRA-Cxx R01 en

Operating conditions

Load limits

The load limits depend on several factors (e.g. dimensions, design

and material of thermowell). The compact sensor TRA-C10 can

handle pressure up to 16 bar / 232 psia while TRA-C20 and TRA-C30

can handle pressures up to 40 bar / 580 psia. This applies only if the

sensor is mounted properly.

Temperature

Temperature

Temperature

Temperature

Process temperature

Maximum -50...+200°C / -50...+392°F, depending on design and

material.

Pre-configured temperature

ranges

-50...+150°C / -58...+302°F or 0...+100°C / 32... +212°F

(Other ranges available on request)

Ambient temperature

Maximum -40...+85°C / -40...+185°F, depending on media

temperature, sensor design and material.

Storage temperature

-40...+70°C / -40...+158°F at 40...60% relative humidity.

Other conditions

Other conditions

Other conditions

Other conditions

Protection categories

Following protection categories can be provided if the connection

cable is properly connected:

OPTITEMP TRA-C10, TRA-C30: IP67

OPTITEMP TRA-C20: IP65

Shock resistance:

Acc. to IEC60068-2-31, test Ec

Vibrations resistance:

Acc. to IEC60068-2-6, test Fc, 10-2000Hz, 4g

Installation conditions

Insertion angle

90° to the flow, directly against or diagonally against the flow.

Insertion length

Recommendable: 10...15 x thermowell diameter for gas,

5...10 x thermowell diameter for liquids (short insertion lengths may

impair measuring accuracy).

Materials

Thermowells (including process

connection)

The metal materials of the thermowells, neck tubes, flanges and

threads as process connections are always identical:

Standard:

Standard:

Standard:

Standard:

1.4404 / AISI 316 L (X2CrNiMo 17-12-2)

Process connections

Threaded thermometer

Gastight thread acc. to ISO 228/1: G½, other sizes on request.

Attachment using welding sleeves. Starting at a wall thickness of

20 mm / 0.8", it is possible to screw in directly with thread in the pipe

Flange thermometer

Acc. to ISO 2852: DN25/38, other sizes on request.

.book Page 34 Friday, February 7, 2014 9:17 AM