KROHNE ALTOFLUX IFM x080 K ATEX EN User Manual
Page 6
ALTOFLUX IFM x080 K-EEx and ALTOFLUX IFM x080 K / i-EEx
6
1.2
PROFIFLUX IFM 5080 K … EEx
The Profiflux IFM 5080 K/…-EEx magnetic-inductive compact flowmeter is in accordance with the
European Directive 94/9/EC (ATEX 100a) and approved for hazardous classified locations of
Zone 1 and 2 under EC-type Examination Certificate number:
KEMA 01 ATEX 2262 X
The compact flowmeter is available in two types, namely:
• IFM 5080 K-EEx regular explosion protected version;
• IFM 5080 K/i-EEx, MODIS version. This type has intrinsical safe signal output circuits, which
are provided by two on the IFC 090i-EEx electronics unit installed MODIS modules (see
Sect. 1.5 for details.
The regular IFM 5080 K-EEx compact flowmeter as well as the so-called MODIS version type
IFM 5080 K/i-EEx is designed for ambient temperatures (T
a
) of -20°C up to +60°C. The allowed
process liquid temperature is limited by the maximum occuring surface temperature of system
components in the combustible atmosphere that (possibly) surrounds the apparatus, determined
by the temperature class of the atmosphere, see table 1 below.
Temperature classification
Maximum process liquid temperature
Temperature
Class
Max. surface
Temperature
T
a
≤ 40°C
T
a
≤ 50°C
T
a
≤60 °C
T6
T80°C
60°C
55°C
not allowed
T5
T95°C
75°C
75°C
70°C
T4
T130°C
115°C
115°C
75°C
T3
T165°C
155°C
135°C
75°C
The IFM 5080 K/…-EEx compact flowmeter consists of the IFC 090/…-EEx signal converter unit,
which is screwed on top of the IFS 5000-EEx primary head (i.e. measuring unit). The compact
flowmeter is marked with the following code below:
II 2 GD EEx dme [ib] IIC T6…T3
In case of the MODIS version IFM 5080 K/i-EEx, the electronics unit of type IFC 090i-EEx is
provided with protective modules, which provide intrinsical safe output signals of category "ia".
The flowmeter is then marked with the code
II 2 GD EEx dme [ib] [ia] IIC T6…T3
For details see the EC-type Examination Certificate in Sect. 8.2 at the end of these additional
instructions.