KROHNE ALTOFLUX IFM x080 K ATEX EN User Manual

Page 16

background image

ALTOFLUX IFM x080 K-EEx and ALTOFLUX IFM x080 K / i-EEx

16

Terminal compartment MODIS version IFC 090i-EEx

Sticker with handling information
for insulating cover plate

Sticker indicating the intrinsically
safe signal in-/output circuits

Insulating cover plate


The PE (or FE) conductor must be connected to the press-fitted M5 clamp terminal marked inside
the terminal compartment. This conductor must be guided through the rectangular opening in the
metal dividing plate that separates the non-intrinsically safe power supply terminals from the
intrinsically safe signal in-/output terminals.

2.6

Connection diagrams MODIS

Sect. 2.3 shows the block diagram of the EEx electromagnetic compact flowmeter. The power
supply (terminals 1L

½, 0L½) is connected via cable B. The PE terminal must be connected to the

protective ground conductor of the mains supply.

The IFC 090i-EEx electronics unit is provided with intrinsically safe signal in-/output circuits due to
the installed pair of MODIS modules in accordance with the table below.