KROHNE ALTOFLUX IFM x080 K ATEX EN User Manual
Page 11
ALTOFLUX IFM x080 K-EEx and ALTOFLUX IFM x080 K / i-EEx
11
1.5
Electronics unit
This electromagnetic compact flowmeter can be equipped with the regular IFC 090-EEx or with
the IFC 090i-EEx electronics unit with intrinsically safe signal outputs (i.e. MODIS version). This
version is descripted in the following.
Regular IFC090-EEx electronics unit
The IFC 090-EEx is used in the regular IFM 4080 K-EEx and can be equipped with one of the
following power supplies (depends on the area of application).
Electrical data for power supply
Power supply
Terminal
Function
Electrical data
AC-versions
L
N
PE
Live
Neutral
Protective Ground
U
n
= 100/115/200/230 V
ac
-15/+10%
P
n
= approx. 10 VA, U
m
= 253 V
AC/DC-version
1L
½
0L
½
FE
Live
Neutral
Functional Ground
U
n
= 24 V ac/dc
AC: -15/+10%, P
n
=10 VA
DC: -25/+30%, P
n
=8 W
U
m
= 253 V
The IFC 090-EEx electronics unit is equipped with the following in-/output circuits. Terminals B1,
B
⊥
and B2 can be configured as status or pulse outputs or as control inputs via the software. See
the table below for the electrical data of these in-/output circuits.
Electrical data of in-/output circuits
Terminals
Description
Nominal voltage
Maximum current
I+, I
Current output
15 V
22 mA
B1, B
⊥, B2
Pulse, status, control in-/outputs
32 V
150 mA
IFC 090i-EEx unit with MODIS modules
The IFC 090i-EEx electronics unit is equipped with a pair of MODIS-modules. It is equipped with
one of the following power supplies.
Electrical data of IFC 090i-EEx electronics unit
Power supply
Terminals Function
Electrical data
AC-version
L
N
PE
Live
Neutral
Protective Ground
U
n
= 100…230 Vac –15%/+10%
P
n
= 15 VA, U
m
= 253 V
AC/DC-version
1L
½
0L
½
FE
Live
Neutral
Functional Ground
U
n
= 24 Vac/dc
AC:
-15%/+10% or 20.4…26.4 Vac
DC:
-25%/+30% or 18…32 Vdc
P
n
= 10 W, U
m
= 253 V
NOTE:
The mains fuses for both electronics units are listed in Sect. 6 of this manual.