4 reset totalizer and cancel error messages, 5 examples of setting the signal converter – KROHNE IFC 210 E Short EN User Manual
Page 27
Part B IFC 210 Signal converter Sect. 4.4
05/2002
IFC 210 E 27
4.4
Reset totalizer and cancel error messages
Cancel error messages in RESET / QUIT menu
Key Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
Code 2
- -
Key in entry code 2 for Reset / Quit
menu:
↑
→
↑
→
Error Quit ....
Menu for error acknowledgement
...
No
Do
not delete error messages,
press
↵
twice = return to measuring mode.
↑
...
Yes
Delete error messages
↵
Reset totalizer …
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
Reset totalizer(s) in RESET / QUIT menu
Key
Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
Code 2
- -
Key in entry code 2 for Reset / Quit
menu:
↑
→
↑
→
Error Quit ....
Menu for error acknowledgement
↵
Reset totalizer …
Menu for resetting totalizer
→
...
No
Do
not delete error messages,
press
↵
twice = return to measuring mode.
↑
... Yes
Reset totalizer
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
4.5
Examples of setting the signal converter
As an example the cursor, flashing part of display, is shown below in bold type.
•
Change measuring range of current output and value for error messages (Fct. 1.05):
•
Change measuring range from 04-20 mA to 00-20 mA
•
Change value for error messages from 0 mA to 22 mA
Key Display
Description
→
If “Yes” set under Fct. 3.04 Entry Code, key in the
9-keystroke Code 1 now:
→
→
→
↑
↑
↑
↵
↵
↵
Fct. 1.00
Operation
→
Fct. 1.01 Full
Scale
4
×
↑
Fct. 1.05
Currentoutput I
→
Function
→
↵
Range
I
→
04-20
mA
Old current range
2
×
↑
00-20
mA
New current range
2
×
↵
I
Error
→
0
mA
Old value for error messages
↑
22 mA
New value for error messages
↵
Fct. 1.05
Currentoutput I
↵
Fct. 1.00
Operation
↵
Store Yes
↵
- - - - - - -
- - - - - / - - -
Measuring range with new data for the current output