Part B IFC 210 Signal converter

22

IFC 210 E

05/2002

Fct.

Display- Texts

Description and settings

Messages

Additional displays wanted in measuring mode?

• no

• yes (overlay additional messages)

Press

key to transfer to subfunction “Trend”.

Trend

Set graphic display

Selection:

• Average

(show average values over time base)

• Min. & Max.

(min./max. values over time base)

• Every Value

(all values numbered consecutively)

Press

key to transfer to subfunction “Updating”.

When

“every value“

selected, transfer to subfunction

“Scaling“.

Updating

Update measured values in graphic display

(this not shown when “every value“, see above, selected)

Selection:

• 0.1 Sec.

• 0.2 Sec.

• 0.5 Sec.

• 1 Sec.

• 2 Sec.

• 5 Sek.

• 1 Min.

• 2 Min.

• 5 Min.

Press

key to transfer to subfunction “Scaling”.

Scaling

Set scaling of graphic display

Selection:

• 0% ... 100%

• - 25% ... 100%

• 0% ... 50%

• 50% ... 100%

• -100% ... 0%

• 25% ... -100%

• - 50% ... 0%

• -100% ... - 50%

• Auto

Press

key to return to Fct. 1.04 DISPLAY.

1.05

Current output I

Set Current output I

Function

Set

Function

• Off

(switched off)

• 1 Direction

(1 flow direction)

• 2 Directions

(forward/reverse flow, F/R flow measurement)

Press

key to transfer to subfunction “Range I”.

Range I

Set measuring range (I

0%

... I

100%

)

• 0 - 20 mA

• 4 - 20 mA (fixed ranges))

• mA (user-defined range)

• Range:

I

0%

- I

100%

(Value I

0%

< I

100%

!)

0 - 16 mA 4 - 20 mA

Press

key to transfer to number setting.

Press

key to transfer to subfunction “I Max”.

I Max

Set I Max

Selection:

• 20,5 mA

• 22 mA

Press

key to transfer to subfunction “I Error”.

I Error

Set current for error identification (I

Error

)

• 22 mA

• 0.0 mA (... < I

0%

, variable only, see above ifI

0%

> 1 mA)

Press

key to transfer to number setting.

Press

key to transfer to subfunction “Range I”.

Reverse Range

Set full-scale range for reverse flow

(appears only when 2 Directions selected, see above)

setting range:

005 - 150% of Q

100%

(different value for reverse flow)

Press

key to transfer to number setting.

Press

key to return to Fct. 1.05 “Current output I”.