KROHNE IFC 210 E Short EN User Manual
Page 22
Sect. 4.2
Part B IFC 210 Signal converter
22
IFC 210 E
05/2002
Fct.
Display- Texts
Description and settings
Messages
Additional displays wanted in measuring mode?
• no
• yes (overlay additional messages)
Press
↵
key to transfer to subfunction “Trend”.
Trend
Set graphic display
Selection:
• Average
(show average values over time base)
• Min. & Max.
(min./max. values over time base)
• Every Value
(all values numbered consecutively)
Press
↵
key to transfer to subfunction “Updating”.
When
“every value“
selected, transfer to subfunction
“Scaling“.
Updating
Update measured values in graphic display
(this not shown when “every value“, see above, selected)
Selection:
• 0.1 Sec.
• 0.2 Sec.
• 0.5 Sec.
• 1 Sec.
• 2 Sec.
• 5 Sek.
• 1 Min.
• 2 Min.
• 5 Min.
Press
↵
key to transfer to subfunction “Scaling”.
Scaling
Set scaling of graphic display
Selection:
• 0% ... 100%
• - 25% ... 100%
• 0% ... 50%
• 50% ... 100%
• -100% ... 0%
• 25% ... -100%
• - 50% ... 0%
• -100% ... - 50%
• Auto
Press
↵
key to return to Fct. 1.04 DISPLAY.
1.05
Current output I
Set Current output I
Function
Set
Function
• Off
(switched off)
• 1 Direction
(1 flow direction)
• 2 Directions
(forward/reverse flow, F/R flow measurement)
Press
↵
key to transfer to subfunction “Range I”.
Range I
Set measuring range (I
0%
... I
100%
)
• 0 - 20 mA
• 4 - 20 mA (fixed ranges))
• mA (user-defined range)
• Range:
I
0%
- I
100%
(Value I
0%
< I
100%
!)
0 - 16 mA 4 - 20 mA
Press
→
key to transfer to number setting.
Press
↵
key to transfer to subfunction “I Max”.
I Max
Set I Max
Selection:
• 20,5 mA
• 22 mA
Press
↵
key to transfer to subfunction “I Error”.
I Error
Set current for error identification (I
Error
)
• 22 mA
• 0.0 mA (... < I
0%
, variable only, see above ifI
0%
> 1 mA)
Press
→
key to transfer to number setting.
Press
↵
key to transfer to subfunction “Range I”.
Reverse Range
Set full-scale range for reverse flow
(appears only when 2 Directions selected, see above)
setting range:
005 - 150% of Q
100%
(different value for reverse flow)
Press
→
key to transfer to number setting.
Press
↵
key to return to Fct. 1.05 “Current output I”.