Part B IFC 210 Signal converter

26

IFC 210 E

05/2002

Fct.

Display-Texts

Description and settings

3.07

Hardware

Assign outputs and inputs to terminals B1 and B2

Terminal B1

Define function of terminal B1 (valid for Fct. 1.07)

• Statusoutput • Controlinput (

Select with

key.

)

Press

key to transfer to subfunction “Terminal B2”.

Terminal B2

Define function of terminal B2 (valid for Fct. 1.08)

• Statusoutput • Controlinput (

Select with

key.

)

Press

key to transfer to subfunction “Fieldcurrent”.

Fieldcurrent

Define field current supply

• Intern

• Extern (> DN 1600 / > 64“ with power driver)

Press

key to return to Fct. 3.07 “Hardware”.

3.08

Location

Set measuring point tag

Factory

setting: ALTOMETER

Characters assignable to each place:

• A-Z, a-z, 0-9 or „-“ (=blank character).)

Press

key to return to Fct. 3.08 “Location”.

3.09

Communication

Set comminucation interface

• Off

(switched off))

HART (HART

®

-interface switched on))

• KROHNE

(KROHNE RS 485-interface switched on),

(only provided if daughter board installed (option).)

• Address:

„HART“ 00-15 / „KROHNE“ 000-239

• Baud rate: 1200, 2400, 4800, 9600 or 19200

(appears with selection “KROHNE” only)“)

Press

key to return to Fct. 3.09 “Communication”.


4.3

Error messages in measuring mode

The following list gives all errors that can occur during process flow measurement. Errors shown
in display when “Yes” set in Fct. 1.04 Display, subfunction “Messages.”.

Error messages

Description of error

Error clearance

Line Int.

Power failure Note:
no counting during power failure

Cancel error in Reset-Quit menu, see
Sect. 4.6.
Reset totalizer if necessary.

Overflow I

Current output overranged
(Flow > I Max)

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

Overflow P

Pulse output overranged.
(Flow > I Max)

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

Totalizer

Totalizer has been reset.

Cancel error message in
Reset/Quit. menu, see Sect. 4.6.

ADC

Analog / digital
converter overranged

Error message deleted automatically
after elimination of cause.

Fatal Error

Fatal error, all outputs set
to “min. values“

Please consult factory.

Empty Pipe

Pipe has run dry.
This message appears only when
the “empty pipe identifier”
option is installed
and the function
is switched on under Fct. 3.06
Application, submenue ”Empty Pipe”.

Fill pipe.