3 error messages in measuring mode – KROHNE IFC 210 E Short EN User Manual
Page 26
Sect. 4.3
Part B IFC 210 Signal converter
26
IFC 210 E
05/2002
Fct.
Display-Texts
Description and settings
3.07
Hardware
Assign outputs and inputs to terminals B1 and B2
Terminal B1
Define function of terminal B1 (valid for Fct. 1.07)
• Statusoutput • Controlinput (
Select with
↑
key.
)
Press
↵
key to transfer to subfunction “Terminal B2”.
Terminal B2
Define function of terminal B2 (valid for Fct. 1.08)
• Statusoutput • Controlinput (
Select with
↑
key.
)
Press
↵
key to transfer to subfunction “Fieldcurrent”.
Fieldcurrent
Define field current supply
• Intern
• Extern (> DN 1600 / > 64“ with power driver)
Press
↵
key to return to Fct. 3.07 “Hardware”.
3.08
Location
Set measuring point tag
Factory
setting: ALTOMETER
Characters assignable to each place:
• A-Z, a-z, 0-9 or „-“ (=blank character).)
Press
↵
key to return to Fct. 3.08 “Location”.
3.09
Communication
Set comminucation interface
• Off
(switched off))
•
HART (HART
®
-interface switched on))
• KROHNE
(KROHNE RS 485-interface switched on),
(only provided if daughter board installed (option).)
• Address:
„HART“ 00-15 / „KROHNE“ 000-239
• Baud rate: 1200, 2400, 4800, 9600 or 19200
(appears with selection “KROHNE” only)“)
Press
↵
key to return to Fct. 3.09 “Communication”.
4.3
Error messages in measuring mode
The following list gives all errors that can occur during process flow measurement. Errors shown
in display when “Yes” set in Fct. 1.04 Display, subfunction “Messages.”.
Error messages
Description of error
Error clearance
Line Int.
Power failure Note:
no counting during power failure
Cancel error in Reset-Quit menu, see
Sect. 4.6.
Reset totalizer if necessary.
Overflow I
Current output overranged
(Flow > I Max)
Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.
Overflow P
Pulse output overranged.
(Flow > I Max)
Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.
Totalizer
Totalizer has been reset.
Cancel error message in
Reset/Quit. menu, see Sect. 4.6.
ADC
Analog / digital
converter overranged
Error message deleted automatically
after elimination of cause.
Fatal Error
Fatal error, all outputs set
to “min. values“
Please consult factory.
Empty Pipe
Pipe has run dry.
This message appears only when
the “empty pipe identifier”
option is installed and the function
is switched on under Fct. 3.06
Application, submenue ”Empty Pipe”.
Fill pipe.