2 table of settable functions – KROHNE IFC 210 E Short EN User Manual

Page 21

background image

Part B IFC 210 Signal converter Sect. 4.2

05/2002

IFC 210 E 21

4.2

Table of settable functions

Fct.

Display- Texts

Description and settings

1.00

OPERATION

Operations menu

1.01

FULL SCALE

Full-scale range for flowrate Q

100%

Select unit

m

3

/hr

Liter/Sec

US.Gal/min

user unit, factory set is Liter/hr or US MGal/day (see Fct. 3.05)

Press

key to transfer to number setting.

Setting

ranges:

The ranges are dependent on the meter size (DN) and the

flow velocity (v):

Q

min

= DN

2

×

v

min

Q

max

= DN

2

×

v

max

Nom. dia. /meter size

v

min

= 0,3 m/s (1 ft/s) v

max

= 12 m/s (40 ft/s)

DN 2.5–1600 /

1

/

10

– 64:

0.0053 –

86 859 m

3

/hr

0.0237 –

401 080 US Gal/min

Press

key to return to Fct. 1.01 FULL SCALE.

VALUE P

Pulse value has been changed.

With the old pulse values the output frequency (F)

would have been exceeded or not reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

1.02 TIMECONST.

Time

constant

Select::

• ALL (applies to display and all outputs)

• ONLY Current output (only display, current and status outputs)

Press

key to transfer to number setting.

Range:

• 0.2 – 99.9 Sec

Press

key to return to Fct. 1.02 TIMECONST.

1.03

L.F.CUTOFF

Low-flow cutoff (SMU)

• OFF (fixed values: ON = 0.1% / OFF = 0.2%)

PERCENT (variable values)

ON

OFF

1 – 19%

2 – 20%

Press

key to transfer to number setting.

Note: Cutoff off value must be greater than cutoff on value.

Press

key to return to Fct. 1.03 L.F. CUTOFF.

1.04 DISPLAY

Display

functions

Contrast

Set display contrast

• range from + 15 (high contrast) to - 15 (low contrast)

Press

key to transfer to subfunction “Flow”.

Flow

Set format for flow rate display

• ####

• ### . #

• ## . ##

• # . ####

• Auto

Press key

to move to unit selection.

Selection unit: •

m

3

/hr

• Liter/Sec

• US Gal/min

user unit, factory set is Liter/hr or US MGal/day (see Fct. 3.05)

Press

key to transfer to subfunction “Totalizer”.

Totalizer

Set format for display of volume

Select decimal positions:

• #######

• ###### . #

• ##### . ##

• #### . ###

• ### . ####

• ## . #####

• # . ######

• Auto

Press

key to move to unit selection.

m

3

• Liter

• US Gal

user unit, factory set is Liter or US MGal (see Fct. 3.05)

Press

key to transfer to subfunction “Messages”.

π

4

π