Part B IFC 210 Signal converter Sect. 4.2
05/2002
IFC 210 E 21
4.2
Table of settable functions
Fct.
Display- Texts
Description and settings
1.00
OPERATION
Operations menu
1.01
FULL SCALE
Full-scale range for flowrate Q
100%
Select unit
•
m
3
/hr
•
Liter/Sec
•
US.Gal/min
•
user unit, factory set is Liter/hr or US MGal/day (see Fct. 3.05)
Press
→
key to transfer to number setting.
Setting
ranges:
The ranges are dependent on the meter size (DN) and the
flow velocity (v):
Q
min
= DN
2
×
v
min
Q
max
= DN
2
×
v
max
Nom. dia. /meter size
v
min
= 0,3 m/s (1 ft/s) v
max
= 12 m/s (40 ft/s)
•
DN 2.5–1600 /
1
/
10
– 64:
0.0053 –
86 859 m
3
/hr
0.0237 –
401 080 US Gal/min
Press
↵
key to return to Fct. 1.01 FULL SCALE.
VALUE P
Pulse value has been changed.
With the old pulse values the output frequency (F)
would have been exceeded or not reached.
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Check new values!
1.02 TIMECONST.
Time
constant
Select::
• ALL (applies to display and all outputs)
• ONLY Current output (only display, current and status outputs)
Press
↵
key to transfer to number setting.
Range:
• 0.2 – 99.9 Sec
Press
↵
key to return to Fct. 1.02 TIMECONST.
1.03
L.F.CUTOFF
Low-flow cutoff (SMU)
• OFF (fixed values: ON = 0.1% / OFF = 0.2%)
PERCENT (variable values)
ON
OFF
1 – 19%
2 – 20%
Press
→
key to transfer to number setting.
Note: Cutoff off value must be greater than cutoff on value.
Press
↵
key to return to Fct. 1.03 L.F. CUTOFF.
1.04 DISPLAY
Display
functions
Contrast
Set display contrast
• range from + 15 (high contrast) to - 15 (low contrast)
Press
↵
key to transfer to subfunction “Flow”.
Flow
Set format for flow rate display
• ####
• ### . #
• ## . ##
• # . ####
• Auto
Press key
→
to move to unit selection.
Selection unit: •
m
3
/hr
• Liter/Sec
• US Gal/min
user unit, factory set is Liter/hr or US MGal/day (see Fct. 3.05)
Press
↵
key to transfer to subfunction “Totalizer”.
Totalizer
Set format for display of volume
Select decimal positions:
• #######
• ###### . #
• ##### . ##
• #### . ###
• ### . ####
• ## . #####
• # . ######
• Auto
Press
→
key to move to unit selection.
•
m
3
• Liter
• US Gal
user unit, factory set is Liter or US MGal (see Fct. 3.05)
Press
↵
key to transfer to subfunction “Messages”.