Sect. 2.3

Part A System installation and start-up

14

IFC 210 E

05/2002

2.3

Status outputs B1 and B2

The status outputs are galvanically isolated from the current output and all input circuits.

Setting data and functions can note down in Section 3.3.
Please also refer to Sect. 3.2 Factory settings.

Typical status outputs B1 and/or B2

B1/B2 status outputs B1/B2



B

chassis ground,

binary outputs and inputs

All operating data and functions can be set,
see Sect. 4 and 5.9, Fct 1.07 and/or 1.08 for operator control


The status outputs can be operated in the active or passive mode.
Active mode:

The current output is the internal voltage source.

Passive mode:

External DC or AC voltage source required.


Characteristics of the status outputs

Switch open

Switch closed

OFF (switched off)

no function

ON (e.g. operation indicator)

power OFF

Hilfsenergie EIN

SIGN I (F/R mode)

Forward flow

Reverse flow

SIGN P (F/R mode)

Forward flow

Reverse flow

TRIP POINT (limit switch)

inactive

active

AUTO RANGE (automatic range
change)

high range

low range

OVERFLOW I (I overranged)

current output OK

current output overranged

OVERFLOW. P (P overranged)

pulse output OK

pulse output overranged

SMU I (low-flow cutoff active)

Inactive

active

SMU P (low-flow cutoff active)

Inactive

active

Inverse B1 (switches B2 inversely to B1)

B2 open, B1 closed

B2 closed, B1 open

ALL. ERROR (all errors)

errors no

error

FATAL.ERROR

(fatal errors only)

errors no

error

EMPTY PIPE (option)

when measuring tube is empty

when measuring tube is full

Connection diagrams, see Sect. 2.5: diagrams

g

h

k

l