2.3 Connection diagrams for
outputs

DC voltage, external power source (U

ext

), note connection polarity

External voltage source (U

ext

), DC or AC voltage, connection polarity arbitrary

Current output

Current output

I

active

I

passive

Active mode

The current output supplies the
power for operation of the
outputs.

Passive mode

External power source
required for operation of the
outputs.

I

= 0/4 - 20 mA

I

= 0/4 - 20 mA

U

ext

15...20 V DC

20...32 V DC

R

i

≤ 500 Ω

R

i

0...500 Ω 250...750

Pulse output P

passive

Pulse

output

P

active

for electronic or

(and current output Iactive) for electronic (EC) totalizers with and

electromechanical

without current output I

totalizers










U

ext

≤ 32 V DC/ ≤ 24 V AC U

int

≤ 15 V DC from current output

I

max

≤ 150 mA

Operation with current output:

(incl. status output)

I

max

≤ 3 mA

Operation

without current output:

I

max

≤ 23 mA

15

V

R ≤

I

max

Status output S

passive

Status output S

active

with and without current output I

U

int

≤ 15 V DC

from current output

I

max

≤ 3 mA

Operation with current output

I

max

≤ 23 mA

Operation with current output

U

ext

≤ 32 V DC/ ≤ 24 V AC

I

max

≤ 150 mA

(incl. status output)

U

ext

U

ext

Load

Load