16 IFC

010

4.3 Error messages in
measuring mode

The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES” set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.”

Error messages

Description of error

Error clearance

LINE INT.

Power failure
Note: no counting during power failure

Cancel error in RESET/QUIT. Menu
Reset totalizer if necessary.

CUR. OUTP. I

Current output overranged.

Check and if necessary correct
instrument parameters. After elimination
of cause, error message deleted
automatically.

PULSOUTP. P

Pulse output overranged.
Note: totalizer deviation possible.

Check and if necessary correct
instrument parameters. After elimination
of cause, error message deletes
automatically.

ADC

Analog / digital
converter overranged

Error message deletes automatically after
elimination of cause.

FATAL. ERROR

Fatal error, all outputs set to “min. values”

Please consult factory.

TOTALIZER

Totalizer has been reset

Cancel error message in
RESET/QUIT menu.

EMPTY PIPE

Pipe has run dry. Message appears only
when the “empty pipe identifier” option
is installed
and the function is activated
under Fct. 3.06 APPLICAT., submenu
“EMPTY PIPE”.

Fill pipe.