beautypg.com

KROHNE IFC 010 Converter Short EN User Manual

Page 6

background image


6 IFC

010

Preparation for connection to IFC 010 W signal converter

Length Ferrule

mm inch

a

90 3.60

b

8

0.30

c

25 1.00

d

8

0.30

External shielding of DS signal cable

Wrap stranded drain wire (7) around the metal foil and clamp under the shield terminal in the signal converter
terminal box.

1.3.4 Cable lengths (max.
distance between signal
converter and flow sensor)

Abbreviations and explanatory notes
used in the following tables, diagrams and connection diagrams

C

Field current cable C, with single shielding, type and length see Table

D

High-temperature silicone cable, 3

× 1.5 mm

2

(14 AWG) Cu, with single shielding, max. length 5 m (16 ft)

E

High-temperature silicone cable, 2

× 1.5 mm

2

(14 AWG) Cu, max. length 5 m (16 ft)

L Cable

length

CB

Intermediate connection box required in connection with cables D and E for flow sensors

OPTIFLUX 4000 F, 5000 F and 6000 F in cases where process temperatures exceed 150 °C (302 °F)

Recommended length of signal cable

for magnetic field frequency

≤ 1/6 × power frequency

Flow sensor

Meter size

Signal cable

DN

mm

inch

OPTIFLUX 1000 F 10

– 15

3

/

8

1

/

2

DS

A4

25

– 150 1

– 6

DS A3

AQUAFLUX F

10

– 1000

3

/

8

– 40

DS

A1

OPTIFLUX 4000 F 10

– 150

3

/

8

– 6

DS A2

200

1000

8

40

DS

A1

OPTIFLUX 5000 F 4.5 – 15

1

/

8

1

/

2

DS

A4

25

– 100 1

– 4

DS A2

OPTIFLUX 6000 F 10

– 15

1

/

8

1

/

2

DS

A4

25

– 80 1

– 3

DS A4

Field current cable C: max. length and min. cross-section

Length

Type of cable, single shielding

0 - 150 m

5 - 500 ft

2

× 0.75 mm

2

Cu / 2

× 18 AWG

150 - 300 m

500 - 1000 ft

2

× 1.50 mm

2

Cu / 2

× 14 AWG

Warning:
Instrument must be properly grounded to avoid electrical shock.

DS signal cable
bending radius
≥ 50 mm (≥ 2")

Cable
length

Cable
length

Electrical conductivity of process liquid

Electrical conductivity of process liquid

DS1

DS2

DS3

DS4

DS1

DS2

DS3

DS4