10 supplement, 1 technical data – KROHNE SU 501 EX EN User Manual
Page 31
10 Supplement
10.1 Technical data
General data
Series
module unit with plug-in socket for mounting on
carrier rail 35x7.5 or 35x5 acc. to EN 50022
Dimensions
W = 36 mm (1.42 in), H = 118.5 mm (4.66 in), D
= 134 mm (5.28 in)
Weight
approx. 170 g (6 oz)
Housing material
Noryl SE100, Lexan 920A
Socket material
Noryl SE100, Noryl SE1 GFN3
Screw terminals
max. 1x1.5 mm²
Power supply
Power supply
20 … 250 V AC, 50/60 Hz, 20 … 72 V DC
Power consumption
max. 3 W (3 … 18 VA)
Sensor input
Quantity
1
Data transmission
Analogue
Hysteresis
100 µA
Switching threshold
12 mA
Current limitation
24 mA (permanently short-circuit proof)
Sensor power supply
15 … 18 V DC
Detection line break
<=3.6 mA
Detection shortcircuit
>=21 mA
Connection cable
2-wire
Resistance per conductor
max. 35 Ohm
Relay output
Number, function
1x switching relay (spdt)
Integration time
0.2 … 20 s, directional switching
Mode
A/B switch (A - max. detection or overfill
protection; B - min. detection or dry run
protection)
Contact
1x spdt
Contact material
AgNi, hard gold-plated
Turn-on voltage
min. 10 mV DC, max. 250 V AC, 250 V DC
Switching current
min. 10 µA DC, max. 3 A AC, 1 A DC
Breaking capacitance
max. 500 VA, max. 54 W DC
SU 501 Ex - Signal conditioning instrument
31
Supplement
27953
-EN
-050616