7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Application range
Temperature measurements of solids, liquids and gases in industrial
environment.
Design
Versions
Versions
Versions
Versions
TT 20
In-head transmitters which are intended for installation in a "B connection
head" or larger according to DIN 43729.
Features
Features
Features
Features
Sensor matching
A matching to a calibrated temperature sensor can easily be performed by
entering the sensor deviation in the low and high ends of the measuring
ranges.
PC programmable
Input type and measuring ranges are set from PC.
Full accuracy is provided without any need for calibration.
Configuration without external power.
Edit/Read configuration offline, i.e. without power supply, by just connecting
to a USB port of a PC.
Measuring accuracy
Accuracy & Stability
Basic accuracy is max. of ±0.1% of span.
Temperature influence
Deviation from +20°C / +68°F
Max. of ±0.25°C / 25°C or ±0.25% / 25°C
Max. of ±0.5°F / 50°F or ±0.28% / 50°F
Supply voltage influence
Negligible
Long-term drift
±
0.1% of span per year
Operating conditions
Temperature
Temperature
Temperature
Temperature
In-head transmitter
Operating and storage temperature:
Operating and storage temperature:
Operating and storage temperature:
Operating and storage temperature:
Standard version: -40...+85°C / -40...+185°F
Humidity
0...100% RH (non-condensing)
Protection category
Protection category
Protection category
Protection category
In-head transmitter
Protection housing: IP65
Protection terminals: IP10
Installation conditions
Mounting
In-head transmitter: DIN B-head or larger, DIN-rail (with adapter)
For detailed information refer to
Installation on page 11.
Weight
In-head transmitter: 32 g / 0.07 lb
Dimensions
For detailed information refer to
Dimensions on page 25.
.book Page 22 Friday, October 14, 2011 11:22 AM