7.2 Technical data tables

7.2.1 Technical data TR 100

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).

Thin film Pt100 (TF)

Wire-wound Pt100 (WW)

Measuring system

Measuring principle

Resistance thermometer measuring insert acc. to DIN 43735-1.

Type of sensor

Pt100 acc. to DIN EN 60751.

Measuring accuracy

The measuring accuracy varies with the temperature and is described in terms of a tolerance class. Values for the

individual tolerance classes are as follows:

Tolerance class A

±

(0.15 + 0.002 x ItI) °C

Tolerance class B

±

(0.3 + 0.005 x ItI) °C

1/3 tolerance class B

±

1/3 (0.3 + 0.005 x ItI) °C

Ø 3 mm / 0.12"

Tolerance class A

(-50...+300°C / -58...+572°F)

Tolerance class A

(-200...+600°C / -328...+1112°F)

Tolerance class B

(-70...+500°C / -94...+932°F)

1/3 tolerance class B at 0...150°C /

+32...302°F, otherwise tolerance class A

Ø 6 mm / 0.24"

Tolerance class A

(-50...+300°C / -58...+572°F)

Tolerance class A

(-200...+600°C / -328...+1112°F)

Tolerance class B

(-70...+500°C / -94...+932°F)

Tolerance class B

(-50...+600°C / -50...+1112°F),

shock resistant

1/3 tolerance class B at 0°C / +32°F,

otherwise tolerance class A

Calibration of the measuring

inserts

Under normal operating conditions we recommend annual recalibration.

Insulation resistance R

iso

The insulation resistance was measured with 250 VDC (measuring insert: 3 mm) or 500 VDC (measuring insert: 6 mm).

20°C ± 15°C / 68°F ± 27°F

> 1000 MΩ

100°C ± 15°C / 212°F ± 27°F

> 500 MΩ

500°C ± 15°C / 932°F ± 27°F

> 50 MΩ

.book Page 24 Thursday, October 8, 2009 10:26 AM