beautypg.com

Electrical connections, 5 wiring variants – KROHNE OPTITEMP TR 100-TC 100 EN User Manual

Page 16

background image

4

ELECTRICAL CONNECTIONS

16

OPTITEMP TR/TC 100

www.krohne.com

10/2009 - 4000630002 - MA OPTITEMP TR/TC 100 R02 en

4.5 Wiring variants

4.5.1 Pt100 measuring inserts

Resistance thermometers are connected acc. to DIN EN 60751 in three different wiring variants.
Of the three, the 3-wire connection is the most frequently used in the industry:

Applications and features

• The 2-wire connection is used with short cable connections and minimal requirements for

accuracy.

• The 3-wire connection is the industry standard and largely eliminates the effect of supply

resistance.

• The 4-wire connection completely eliminates the distorting influences of the supply

resistance and guarantees minimal measuring uncertainty.

TR 100: Wiring depending on sheath diameter

TR 100: wiring

Figure 4-1: Double (top row) and single version (bottom row). From left to right: 2, 3 and 4-wire connection.

1 white
2 red

Diameter [mm] / [inches]

single Pt100

Double Pt100

3 / 0.12

2-, 3- or 4-wire connection

2-wire connection

6 / 0.24

2-, 3- or 4-wire connection

2-, 3- or 4-wire connection

.book Page 16 Thursday, October 8, 2009 10:26 AM